40 MS-70 / MX-50 / MF-50 / ML-50
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
2. Press the PROGRAM key and press the or
key to select a program number.
3. Press the ENTER key to use the number.
4. The analyzer displays end and returns to the
weighing mode.
Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
Select an Analysis Mode
5. Press the SELECT key to select a symbol, it
will blink. Select U-t of the timer mode with
the or key. (Press these keys to select it)
Set the Heating Pattern
6. Select drying temperature with the SELECT key.
Select step drying of the heating pattern
with the PROGRAM key.
Set Drying Temperature 1
7. Set 120°C with the or key.
Set Time 1
8. Select time 1 with the SELECT key.
Set 5.0 minutes with the or key.
Set Drying Temperature 2
9. Select Drying Temperature 2 with the SELECT key.
Set 160°C with the or key.
Set Time 2
10. Select time 2 with the
SELECT key.
Set 10.0 minutes with the or key.
PROG 1
0.000g
PROG "6
end
PROG 5
0.000g
Program number
1
"1"2"0°C
"U"-"t
0.000g
1
2
2
"1"6"0°C
"1"0"5°C
"1"0."0 min
Drying Temperature 1
Analysis mode
Drying Temperature 2
Step drying
Time 2
Time 1
"5."0 min